NPGS
Microscope
Considerations
|
Home |
Overview | Questions
& Answers | System Description
| Microscope Considerations
| User List |
Sample Pictures |
Other Resources |
Book List | User
Notices | Send
E-mail
|
Select a topic from the following list or scroll
down the page to see all the information available:
|
Microscope Models that have been used
with NPGS:
Almost any SEM or STEM can be used with NPGS to perform e-beam
lithography. The following is an alphabetical list of the microscopes
that have been used with NPGS. Note that some models require optional external
inputs from the manufacturer. Often, different models of microscopes from
the same manufacturer will have the identical XY interface. Consequently,
models that are not shown, may actually be the same as a model already in
use with NPGS. Please call or send e-mail if you would like to get the
complete NPGS User List which includes names, telephone/fax numbers, e-mail
addresses, and microscopes being used. (The models shown below are indicated
as: SEM=Scanning Electron Microscope, STEM=Scanning Transmission Electron
Microscope, FE-SEM=Schottky (thermal) Field Emission SEM, and cFE-SEM=cold
cathode FE-SEM.)
-
Amray 1200, 1400, 1830 SEMs
-
Amray 1840, 1845, 1850 FE-SEMs
-
Cambridge Instruments S120, S200, S240, S250, S360 SEMs
-
Camscan Series 4 SEM
-
Elionix ERA-8800
-
FEI XL30-W, XL30-LaB6, XL40-W SEM
-
FEI XL30 FEG, XL30 SFEG FE-SEMs
-
FEI XL30 ESEM FEG FE-SEM
-
FEI Inspect S, Quanta SEMs
-
FEI Inspect F, Quanta FEG, FEG ESEM FE-SEMs
-
FEI Sirion, NanoSEM, Magellan, Teneo, Apreo, Quattro FE-SEM
-
FEI Quanta 3D Dual W, Ion Beam Microscope
-
FEI Strata 235/237/620/810/820 Dual FE & Ion Beam Microscope
-
FEI Quanta 3D, Nova Nanolab, Helios, Scios Dual FE & Ion Beam
Microscope
-
Hitachi S510, S570 SEMs
-
Hitachi S2460N SEM
-
Hitachi S2300, S2400, S2500 Delta, S2700 SEMs
-
Hitachi S3000H, S3000N, S3400, S3500N SEMs
-
Hitachi S4000, S4100, S4200, S4500, S4700, S4800 cFE-SEMs
-
Hitachi S4300SE FE-SEM
-
Hitachi SU1510, SU3500, SU3800 SEM
-
Hitachi SU6600, SU70, SU5000 FE-SEMs
-
Hitachi HD2700 STEM
-
Hitachi FB2000A FIB
-
Ionoptika Custom FIB
-
ISI 60 SEM
-
JEOL 1200EX STEM
-
JEOL 820, 840, 845, 848 SEMs
-
JEOL 5400, 5600, 5800, 5900, 5910, 6360, 6460 , 6380, 6480, 6390, 6490, 5700,
6010, 6510, 6610 (also LV versions) SEMs
-
JEOL IT100, IT200, IT300, IT500 , IT500HR SEMs
-
JEOL 6100, 6300, 6400, 6600 SEMs
-
JEOL 840F, 6300F, 6340F, 6400F, 7401F, 7500F cFE-SEMs
-
JEOL 6500F, 7000F, 7001F, 7100F, 7200F, 7600F, 7800F, 7900F, F100,
IT800 FE-SEMs
-
JEOL 4501
-
Leica S440 SEM
-
LEO S430, S438, S440 SEMs
-
LEO 1430, 1430VP SEMs
-
LEO 982, 1525, 1530, 1550 (also VP versions) FE-SEMs
-
Philips 501 SEM
-
Philips EM420 STEM
-
Philips CM-20 STEM
-
Philips XL-20, XL-30 SEMs
-
STS SEMView8000
-
Tescan Vega SEM
-
Tescan Mira FE-SEM
-
Tescan Lyra Dual FE & Ion Beam Microscope
-
Thermo Fisher Scientific Apreo
-
Thermo Fisher Scientific Helios, Scios Dual FE & Ion Beam Microscope
-
Topcon SM 350 SEM
-
Zeiss 940A, 960A SEM
-
Zeiss EVO SEM
-
Zeiss Orion He Ion Microscope
-
Zeiss/LEO Supra, Ultra, Sigma, Merlin, GeminiSEM (also VP versions)
FE-SEMs
-
Zeiss/LEO Auriga, Neon, Crossbeam 1540 Dual FE & Ion Beam Microscope
Notes: Cambridge Instruments became Leica, which merged with Zeiss to become
LEO, which has now changed to Carl Zeiss - Nano Technology Systems; Philips
became FEI, which became Thermo Fisher Scientific (TFS).
|
Overview of Microscope
Considerations: (go to Top)
The following is an overview of the considerations that are important when
choosing a microscope to be used for lithography. (Also see the summary of
microscope features that are useful for lithography on the
"Questions and Answers" page.)
-
Overall Performance- The research or analytical level of microscopes
tend to perform the best for lithography, while the low-cost, entry level
microscopes tend to have more problems. Often an older, high performance
microscope will work better for lithography than a new low-cost model.
-
Beam Current Stability - Requirements depend on pattern linewidths
and total exposure time. Pattern exposure times may range from seconds to
hours. Tungsten and LaB6 filaments have performed very well, while cold
cathode field emission (cFE) filaments tend to drift significantly and have
significant noise. When considering a cFE SEM for lithography, getting
a graph of the beam current vs. time will be very helpful in evaluating the
potential problems it may impose on lithography. NPGS users with cFE SEMs
have reported at least +/-3% noise and typically +/-15% drift per hour, and
in some cases random +/-10% jumps and/or drifts of 20% in 20 minutes have
been observed. In contrast, thermal FE sources are very stable and have low
noise. For example, measurements taken in one NPGS user's lab of the beam
current from a thermal FE source over a 12 hour period showed a max to min
variation of less than 1 pA, while the average value was 129.5 pA. This was
less than 0.1% per hour drift . About 97% of the readings (which were taken
every 5 seconds) showed no noise (to the 0.1 pA resolution of the meter)
and the remaining 3% of the readings had values within 0.4 pA of the local
average. Such stability is ideal for SEM lithography. In contrast, a cold
cathode FE SEM will often have continuous noise spikes/jumps on the order
of +/- 3%, less frequent random jumps on the order of +/- 10%, and drifting
on the order of 10% per hour. The best approach that is known to reduce the
instability of a cFE SEM is to frequently bake the gun.
-
Maximum Beam Current - Tungsten and LaB6 filaments provide more than
enough beam current for most SEM lithography applications. Older thermal
FE sources provide much less beam current than conventional sources, however,
the maximum should be adequate for almost all SEM lithography. Note
that some of the newer thermal FE SEM models can provide high beam currents
(>100 nA). Cold cathode FE sources provide even less beam current than
thermal FE sources, which may impose a limit when trying to write very large
structures quickly by using a large beam current. All sources provide
adequate current for writing fine structures.
-
Resolution - The ultimate resolution of the microscope is not the
most important concern for lithography, although a high
attainable resolution usually indicates a high performance microscope.
Being able to consistently minimize the astigmatism is the more important
issue for most lithography.
-
Accelerating Voltage - In general, a higher beam voltage produces
smaller linewidths. However, by using multilayer resists, surprisingly high
resolution lithography can be done at relatively low (<15 kV) beam energies.
Older high performance SEMs will often go to 40 kV, while new SEMs typically
have a 30 kV maximum voltage. At the same kV, a field emission SEM will typically
produce finer lines than a conventional SEM.
-
Magnification - Any microscope magnification may be used for pattern
writing. The magnification will determine the patterning field of view and
the minimum linewidth. Typically, fine patterns (< 0.1 um) will be written
at a magnification that gives a field of view of approximately 100 um square.
Also, the writing field center will often be offset as the microscope
magnification is changed. NOTE: Any magnification value is the ratio
of the size of the displayed image to the size of the area being viewed.
Historically, all magnification values were displayed relative to a Polaroid,
since that was the only output device of most electron microscopes. However,
on many newer microscopes, the microscope software will allow the magnification
to be displayed relative to a Polaroid, a VideoPrinter, or the PC Display,
consequently, different magnification values will be displayed for the same
sample area, depending on the mode of the microscope software. In summary,
"magnification" is not the nearly universal number as it once was and any
comparisons must account for the mode of the microscope when the value was
displayed. However, many people still talk about magnification values as
though all microscopes still use the Polaroid as a universal reference size.
The difference in the magnification value displayed in the different modes
may range from ~5% to ~300%. For example, a 1,000,000x image on a microscope
in the Display mode with a 21" screen may be about 300,000x when in Polaroid
mode, even though the image itself is identical in both cases.
-
Magnification Transition Values - Every SEM will have special transition
values when changing the magnification which select different amplification
circuits within the microscope. When imaging, an audible 'click' is often
heard and sometimes a momentary 'glitch' is seen on the display, but no
significant change is typically observed when crossing one of these transition
values. However, for lithography, the signal-to-noise within the microscope
will typically be noticeably worse as seen in patterns written when the
magnification is just below a transition value as compared to just above
the transition value. For the best lithography, these transition values,
which often depend both on the accelerating voltage and working distance,
should be identified and the magnification values used for pattern writing
be selected accordingly. The NPGS Manual gives typical transition values
for many SEM models and provides additional information on determining the
optimum magnifications for pattern writing for any microscope. (Note: These
transition values are a completely separate issue from the magnification
modes discussed below.)
-
Magnification Modes - Most SEMs will have a single magnification mode
which ranges from the lowest possible to the highest possible setting. This
single magnification mode is good for lithography, since the optimization
used for high resolution writing is preserved when changing to a lower
magnification for larger structures. However, some field emission SEM models
from Hitachi and JEOL have separate high and low magnification modes where
the ranges of the two modes overlap around a field size of 100x100 um2 such
that larger writing fields required the low mag mode and smaller fields require
the high mag mode. Once the high or low mag mode is set on the microscope,
the only way to reach the field sizes of the other mode is for the microscope
to make substantial changes to the electron optics. This is typically done
by a simple mouse click in the SEM software, however, changing back and forth
will degrade the optimization of the SEM. The nuisance is that many lithography
applications require fine features to be written in the high mag mode and
also large features to be written in the low mag mode, where the SEM optimization
is degraded when switching between the modes. This does not prevent the SEM
from being used for lithography, but it is a negative issue for some
applications.
-
Stage Stability - Basic concerns are backlash and drift. Fine Z motion
is also useful for focusing. If you already have the microscope, it is very
easy to characterize the stage stability simply by going to a high magnification,
moving the stage, and measuring the drift versus time. For a given application,
the requirement depends on the feature size and the writing time. For
example, if a pattern consisting of 0.1 um lines takes 1 minute to write,
a stage drift of 20 nm/minute would probably be acceptable. When a stage
has a significant drift problem, designing the pattern to write the features
requiring careful alignment first can minimize the problem.
-
Electronic Scan Rotation - This is useful for aligning the writing
axes to the sample and/or stage motion, however, on some SEM models the scan
rotation circuit will introduce distortions which make it unsuitable for
use with lithography.
-
Magnetic Shielding - Magnetic shielding may be necessary depending
on the microscope environment. Typically, if no external interference
is observable when the microscope is used for viewing, then pattern writing
should also be unaffected.
-
Faraday Cup - A picoammeter is essential for measuring the beam current
before exposures. Ideally, a Faraday cup will be inserted into the column
or directly above the sample without requiring the sample to be moved, or
it will be mounted on the sample holder. If the Faraday cup is on the stage,
the picoammeter can usually be connected to a "specimen current output" on
the SEM. Normally, the specimen current output will either have a grounding
cap which simply keeps the sample at ground potential or it will be connected
to an alarm circuit which indicates if the sample/stage has touched anything
in the chamber.
-
Beam Blanker - A magnetic or electrostatic beam blanker is suggested,
although one is not required. Using only a slow beam shutter or no blanker
at all is sufficient for many applications. See the
"Questions & Answers" page for more information
about doing lithography without a blanker. When considering a new microscope
it makes sense to purchase a blanker with the microscope. Be sure to check
that the blanker will have electronics that will accept a standard TTL (or
comparable) input and not require a low noise input. Also, a blanker should
be expected to work over the full operating range of the SEM (i.e., from
high current to low current, and from high kV to low kV, etc.) with no
degradation of imaging or blanker performance, unless otherwise noted by
the seller. Be advised that in general, most SEMs are not sold with blankers
and the SEM salespeople are often unfamiliar with the details of the blanker
that their company offers. Note that some blankers with a "Faraday cup" feature
do not give a true reading of the beam that hits the sample!
-
X&Y External Control - The beam position must be controllable
by an external voltage. Most microscopes have an External Beam Interface
which is either built-in or available as an option (see Requirements below).
If the interface is already used by an x-ray or digital imaging system, a
manual switch or relay box will be required.
-
Image Signal Output - An image signal must be available for precision
alignment. If the scope does not have a processed output signal, the amplified
PMT output can be used (see Requirements below).
-
Digital Interface - Typically, the column parameters for a digital
microscopes can be controlled during pattern writing through a serial or
Ethernet port. Free software drivers are available with NPGS for microscopes
from Hitachi, JEOL, Zeiss/LEO/Leica, Tescan, and FEI/Philips. (Please
check on specific models.) The most useful commands through a microscope's
digital interface control the magnification (or FOV), focus (=WD), spot size
(=condenser lens setting or sometimes an aperture size), degauss (=lens clear
or hysteresis removal), x stigmator, and y stigmator. Usually, the same digital
interface is also used to control the XY stage position.
-
Automated Stage - Stages that can be controlled by an external computer
may be used to automatically move the sample during pattern writing. Free
software drivers are available with NPGS for stages from Deben Research,
Hitachi, JEOL, Zeiss/LEO/Leica, FEI/Philips, Kleindiek. (Please check
on specific models.)
-
Other Microscope Accessories - While accessories such as EDS and WDS
are normally not used during lithography, there are lithography applications
that involve gas injection systems (GIS). A free software driver is available
with NPGS for the OmniProbe GIS.
-
Requirements - The microscope must have XY scan control inputs with
a range between +/- 3 and +/- 10 volts with input impedance greater than
2k ohms. Also required is access to an image intensity signal within +/-10
volts. A picoammeter is also required for lithography.
-
Limitations - Typically, microscope limitations are most pronounced
when using a large (> 500 micron) field of view. The limitations include
nonlinearity of the scan coils and position dependent focus and astigmatism
effects. SEM lithography is a demanding application and limitations
can be observed in any commercial electron microscope. Typically, linewidths
on the order of 0.2 micron can be written over a field of 500 microns, while
the finest lines (<0.03 microns) can be written in a field of about 100
microns. Large structures can be written at the largest field of the microscope,
which is usually about 10 mm. Results will vary depending on the microscope
in use.
|
Detailed List of
Requirements/Specifications/Considerations: (go
to Top)
Where appropriate, expected, typical, or preferred responses are
shown below in [ ]'s. While NPGS can be used with the most basic of
manual SEMs, it also has very advanced features that make use of automated
stages and the serial interface to a digital SEM. The items below should
cover the most advanced use of the SEM for lithography.
-
XY Inputs (required)
-Voltage range for full deflection (x & y)? [+/-10v to +/-5v]
-Input impedance (x & y)? [>2k ohms]
-Connector type? [BNC]
-External scan enabled by software or electrical connection? [either is OK]
-Are multiple xy inputs available if NPGS and x-ray system are to be used?
[if not, a special input adapter may will needed if an x-ray mapping system
is present]
-Does internal scan generator turn off when SEM is in external mode? [often
no, preferably yes]
-
Image Signal Output (required)
-Maximum voltage range? [< +/-10v]
-Typical voltage range for black to white image change? [> .3 volts]
-Connector type? [BNC]
-Signal available when SEM is in external mode? [yes]
-
Picoammeter (required)
-Picoammeter available integrated with SEM? [typically no]
-If yes, can it be activated and read through SEM serial interface in pA
or Amps?
-If no, can Keithley 485, 486, or 6485 be connected to Faraday cup on stage?
[yes]
-Connector type for specimen current reading? [BNC or Triax] (This allows
a picoammeter to read the beam current collected by a Faraday cup mounted
on the sample holder. Often, an alarm circuit that indicates if the sample
has hit the final lens assembly will need to be disconnected in order to
read the beam current from the sample holder.)
-
Blanker (optional)
-Typical rise/fall/delay times? [< 1 usec each]
-Maximum repetition rate? [>100 kHz]
-Voltage input (on/off values)? [within +/-5 volts]
-Input impedance? [>50 ohms]
-Input noise threshold? [should not require low noise input]
-Connector type? [BNC]
-
Stage/Sample Holder
-Allows gold resolution standard, Faraday cup, and sample to be mounted
simultaneously? [yes]
-
Automated Stage (optional)
-Serial or Ethernet interface to read and set (x,y) positions? [yes]
-Resolution for reading/setting through serial interface? [1 um or better]
-Accuracy? [typical OEM stages are accurate to 1 to 5 um]
-Backlash correction available? [yes]
-
SEM Interface for Remote Control (optional)
-Interface protocol? [RS232 serial or Ethernet]
-Can read and set fine focus (or working distance) to the same resolution
as through SEM User Interface? [yes]
-Can read and set magnification? [yes]
-Typical image image shift when changing between magnifications? [<5 um
for large mag changes]
-Can set probe current to user defined value (e.g. 10.0 pA)? [typically no]
-Can set fine probe current value (arbitrary units)? [typically yes, allows
indirect setting of current to a specific value, e.g. 10.0 pA]
-If no, can set coarse probe current (or spot size/aperture setting)?
-Can fine/coarse probe current be changed without losing beam optimization?
[fine is typically OK, coarse will usually introduce astigmatism]
-Can access all SEM User Interface commands through serial interface? [list
any exceptions; however the typical parameters changed during lithography
are addressed above]
-
Other SEM Parameters/Considerations
-Minimum magnification when in external mode? [100x or less]
-Drift in probe current? [W is ~1%/hour, LaB6 is ~3%/hour, thermal FE is
<<1%/hour, cold cathode FE is ~10%/hour or worse]
-Noise in probe current? [W, LaB6, & thermal FE ~1% or less, cold cathode
is ~ +/-3% or worse]
-Minimum/Maximum probe currents available? [5 pA to 1 nA]
-Maximum accelerating voltage (kV)? [30 or higher]
-Stability of SEM at maximum voltage (hours)? [3 hours or more]
-Digital image storage to PC formats available/standard? [to floppy, ZIP,
or CD in TIFF, GIF, or JPEG]
-Stage drift (nm/minute)? [<10 nm/min after 4 minutes]
-Stage settling time required after movement to meet drift specification
(seconds)?
-Manual fine Z control of stage available? ["fine" means good enough to use
for focusing at 20,000x]
-Electronic scan rotation available/standard? [must rotate scan when in external
mode without distorting image]
|
Checking the Performance of an
SEM: (go to Top)
-
Microscope Setup - Don't be afraid of the microscope! If you
are in charge of running the microscope, you should be very familiar with
its setup and operation. Don't simply trust that the microscope will always
be in good shape if it is under a service contract. Many times when I have
inspected newly purchased microscopes that were professionally installed
by the various vendors (and often signed off and paid for), I have found
that essential components were installed improperly, incompletely, or just
didn't function, or the microscope had completely unexpected and inexcusable
problems, such as shipping bolts that had not been removed or cables that
were connected incorrectly. Often, a simple visual inspection of the microscope
will find problems such as loose wires or bolts that have been dropped where
they don't belong. Touching the column can identify vibration problems or
thermal problems with coils. You should touch any cables (just avoid any
exposed electrical connections) or hoses that lead to the microscope column
to check for "vibration leaks", for example, stiff water lines from a chiller
can very efficiently transfer vibrations directly to the column. Also, gently
pushing on the column should make it rock back and forth. If not, the vibration
isolation of the column should be checked (this was all that was necessary
to discover that the internal shipping bolts had not been removed during
the 'professional' installation of a new FE SEM, even before seeing the terrible
imaging). If the microscope is on a third party vibration isolation support
structure, be sure that all the support columns are actually floating (two
out of three support structures that I have inspect have been malfunctioning).
Typically, the issues that are discovered in such an inspection will not
prevent the microscope from running, however, they very well may prevent
the microscope from running at its ultimate performance level, which is what
you want for lithography.
-
Room Environment - In many cases, the microscope itself will be performing
as well as it can, however, the room environment (vibration, acoustic noise,
magnetic fields) can degrade the observed performance. For a discussion
on typical concerns, see the online paper:
"Site Selection Considerations For
Scanning and Transmission Electron Microscopes" by Wayne Vogen, Vibration
Engineering Consultants, Inc. See the NPGS page on line
frequency interference, which also lists a source for an inexpensive
gauss meter, which can be extremely useful for locating sources of magnetic
fields. While imaging at high magnification, simple tests such as stomping
your foot, clapping, or even talking can illustrate the sensitivity of your
microscope to external vibrations and acoustic noise.
-
Image Resolution - Take a picture of a gold resolution standard (typically
gold islands on a carbon substrate) at the highest accelerating voltage
(typically 30 or 40 kV), at ~10 pA, and with a field size of about 1
micron. Use the smallest aperture to get the best resolution. Look
for line frequency noise (broad bands of distortion), vibration (choppy
distortion along vertical edges), and overall image sharpness. Line frequency
noise from the environment will be worse when using lower kV or longer working
distances. Comparing to an image from a higher performance SEM or the
gold standard images on the NPGS web site can
be helpful.
-
Focus Stability - The long term stability of the SEM focus can be
checked by doing the following: After focusing on a gold standard, either
blank the beam or move the sample to another area a large distance away and
wait for 10 minutes. Then, return to the original area to check if the sample
is still in focus. For a thorough check, this can be repeated over a period
of 1 to 2 hours.
-
Beam Current Stability - Connect a picoammeter to a Faraday cup in
the SEM and monitor the beam current from the initial SEM startup until the
value stabilizes. Typically, an SEM will take 1/2 to 2 hours to stabilize
to ~1% drift per hour for a W filament, 2% to 3% drift per hour for LaB6,
<1% drift per hour for a Schottky (thermal) FE SEM, or +/-5% to +/-30%
of drift and random jumps for a cold cathode FE SEM. Look for unusual spikes
or abrupt changes in the beam current, which should not be present other
than with a cold cathode FE SEM. (However, there is evidence that frequent
baking of a Hitachi cFE SEM can greatly reduce the beam instability.)
-
Optimum Magnification for Pattern Writing - The typical SEM will have
different circuit elements for different ranges of magnifications and the
low end of each range will provide the best signal to noise for pattern
writing. The signal to noise may change by an order of magnitude
or more when when changing from the magnification at the top of one range
(worst) to the bottom of the next (best). (The signal to noise also
changes when imaging, however, the resolution of images is typically too
low to directly observe the degraded performance, however, the difference
can be dramatic for lithography.) To find the changes in the magnification
ranges, do the following.
-
Optimize the SEM for the conditions to be used for lithography, such as kV,
beam current, and working distance, and change to the lowest possible
magnification.
-
Steadily increase the magnification and watch for any 'glitch' in the image
or offset in the image position (when above ~20,000x) as the magnification
is changed. Also, listen for any audible 'click' which may result when mechanical
relays change from one range to another. Note that lithography is typically
performed with a field size of ~20 microns or larger (up to the maximum possible
field size), so any range changes that occur at magnification values
with smaller fields do not really affect lithography, but can be useful to
check while characterizing the microscope.
-
For analog SEMs, the changes will typically occur at fixed intervals such
as 100x, 1000x, 10,000x, or 250x, 2500x, 25,000x, while for many new PC based
SEMs (from FEI and LEO) the range changes will depend on both the kV and
working distance being used and even on if the magnification is being increased
or decreased (from FEI).
-
Note that for PC based SEMs, the displayed magnification number may refer
to different output dimensions, such as for a Polaroid output, the screen
display, or a videoprinter printout. When the displayed SEM magnification
is calibrated for a videoprinter, the values will be similar to a Polaroid
calibration (since the two output images are nearly the same size), but when
it is calibrated for the SEM display, the magnification value will typically
be about 2 to 2.5 times higher (since the typical screen display size is
about 2 to 2.5 times larger than a Polaroid image).
-
Beam Blanker Response - To be thorough, the blanker should be tested
at low and high beam currents, at low and high accelerating voltages, and
at low and high magnifications.
-
A simple two step test is to put the beam into a Faraday cup on the sample
holder and measure the steady-state beam current and then to measure the
beam while it is being dynamically blanked with a 50% duty cycle. The 50%
duty cycle should produce a beam current reading that is exactly 50% of the
steady state reading. The measured ratio can be checked at different duty
cycles and/or different blanking frequencies.
-
The image itself can also be used to diagnose the blanker. While the beam
is being dynamically blanked, the image should not shift or degrade (other
than the black stripes caused by the blanking). To make the test the most
sensitive, look at a gold resolution standard while at high magnification
so that you can look for small changes in focus and astigmatism that may
be induced by the blanker. You may need to adjust the blanking frequency
and/or the scan rate to obtain black stripes on the screen that are relatively
steady. When using a digital SEM, turn off any averaging of the display image.
-
Obtain an image of a high contrast structure and then change the SEM to line
scan mode. In this mode, a single line will be scanned and the vertical
displacement of the trace on the SEM display will represent the brightness
of the structure being scanned, much like an oscilloscope display. While
observing the line scan mode, blank the beam with the blanker to see of the
entire line scan trace changes to a flat line. If any structure is seen when
the beam is blanked, it indicates that some beam is still hitting the sample,
i.e., the blanker is not blanking completely. To increase the sensitivity
of this test, adjust the brightness and contrast of the SEM so that the blanked
signal is at the bottom of the SEM display and the bright image display is
maximized. Ideally, the blanking ratio for the focused beam will be 1:1,000,000
or more. (In other words, if it takes "X" microseconds to expose a small
dot with the beam on, it will take over "X" seconds to expose the same dot
with the beam blanked.)
-
Check if the settings of the gun tilt/shift or the adjustable aperture affect
the performance of the blanker. If so, be sure that the optimum settings
for blanking are the same as the optimum settings for imaging.
-
By using a dual trace oscilloscope to monitor the input signal to the blanker
(produced by a function generator) and an image output signal from the SEM
(often labeled Video Out), the response of the blanker can be measured. Slow
gun coil blankers will have a response time of 1 to 10 milliseconds, while
a fast blanker will have a response time of 1 microsecond or less. If a blanker
is consistently in the "on" state for less time than indicated by the input
signal, a compensation for lithography can be made in the Pg.sys file (this
typically only applies to ~1MHz magnetic blankers). A 50 kHz blanking signal
is a good frequency to start with for a fast blanker. Note that the beam
should either be stationary or be scanning a featureless area of a sample,
so that there are no changes in the SEM output signal due to sample structure.
Also, note that the response time of the image output signal may be slower
than the blanker response for fast electrostatic blankers, so seeing rise/fall
times of ~0.2 microseconds is not a cause for alarm. Unusual behavior while
the beam is being dynamically blanked (i.e., pulsed by a function generator)
includes a sloped response on the output signal when the beam is on and/or
a delay which varies with beam current or accelerating voltage. The steady
state "beam on" condition will produce a steady (but noisy) trace on the
oscilloscope screen and the steady state "beam off" condition will produce
a different trace on the oscilloscope screen (typically a relatively low
noise signal at a lower voltage than the beam on case). Ideally, when the
beam is dynamically blanked, the oscilloscope trace will cleanly jump from
the "beam on" signal to the "beam off" signal.
-
Unfortunately, the blanker tests listed above cannot guarantee that a blanker
will perform flawlessly for all lithography. However, these tests are relatively
simple and are very quick, so they are extremely helpful for finding obvious
problems with a blanker, while sometimes more subtle problems will only be
observable by writing patterns with fine lines. However, if a problem is
observed in the tests described above, it is then much easier to resolve
the problem, than the case when an entire lithography processing cycle is
involved.
-
Scan Rotation - Scan rotation is a very handy feature to have during
lithography, since it allows the scan field to be rotated without moving
the sample. However, on old SEMs the scan rotation knob my introduce a
significant amount of noise into the XY signal. This can be observed if the
image motion is "jerky" as the knob is changed. Such behavior may indicate
that the scan rotation potentiometer is oxidized. If the pot cannot be easily
accessed for cleaning or replacement, quickly turning the knob back and forth
many times may rub off the oxidation and significantly improve the performance
(the same technique can apply to old stereos where the volume knob introduces
static or causes random jumps in the volume). For new digital SEMs, it has
been observed that they often introduce up to ~3% distortion to the image
XY scaling. While this size distortion will appear in the pattern writing
(unless the NPGS alignment feature is used to scale the pattern to registration
marks), it is typically small enough that the convenience of using the scan
rotation outweighs the aspect ratio problem.
-
Stage Backlash & Stability - While imaging at high magnification,
find a small unique feature on the sample and use this to check the backlash
as the stage controls are move back and forth. A good stage may have backlash
on the order of a micron, while in a bad case, the backlash may be as much
as 30 microns or more. An automated stage controller using stepper motors
attached to the existing stage controls can minimize backlash. Stage stability
can be measured by moving the stage so that the unique feature is centered
on the screen and then measuring how far it moves as a function of time.
Generally, stage drift is not a significant concern during lithography, unless
careful registration between different parts of the pattern is a high priority.
In this case, careful ordering of the pattern elements can minimize the effect
of stage drift. Note that a steady drift in the image can be the result of
either the stage moving or changes in the electron optics. High performance
SEMs typically have an acceptably small drift, while entry level SEMs may
have a very noticeable drift. Tests that can be done are:
-
At high magnification (a field size of 1 micron or smaller), locate a small,
bright bit of something on the sample. Move the stage away and then back
so that the selected structure is nominally centered on the screen. Watch
how much the structure moves following the move. Movement that is initially
fast but then slows may be settling of the stage, which may last from seconds
to minutes. Long term drifting in one direction may be caused by thermal
changes of the microscope column, especially if it has not been operating
long (>30 to 60 minutes). Sudden changes where the image moves and then
returns may be caused by changes in the local magnetic field, which may be
caused by any large metal structures moving nearby, such as steel doors or
elevators.
-
If the stage is automated, center the image on an easily identifiable feature
(with an imaging field size of ~10 microns), then make a series of automated
stage moves away and back to the same location. This will give you an idea
of the reproducibility of the automated stage moves. Start with simple "over
and back" motion, and then try more complicated sequences which return to
the same location. A good stage will give sub-micron positioning accuracy
for a single over and back move and 1 to 2 micron positioning accuracy for
multi-step sequences.
-
If the stage is automated and you have a sample with features of a known
size or displacement (be careful, since some optically defined lithography
patterns may have significant size or positioning errors), program the stage
to move the known distance and compare to the locations on the sample. This
checks the absolute accuracy of the stage moves.
-
Check the backlash of the stage, where backlash is the determined by the
positioning error that results from moving the XY knobs or motors in one
direction versus the opposite direction. For automated stages, there may
be a backlash correction feature that can be disabled. If so, repeat the
automated stage tests described above with the feature on and off.
-
Stage Automation - Most older SEMs have only a manual stage, while
most newer models have an automated stage integrated with the SEM controls.
However, some models have a motorized stage that cannot be
automated, i.e., there is no interface for external control of the
motorized stage. Most manual stages can be automated through the use of third
party stage controllers. Note that while NPGS can be interfaced to any automated
stage, a manual stage is sufficient for many lithography applications. Drivers
for NPGS to control most common automated stages from FEI, LEO, JEOL, Hitachi,
Amray, Deben, and Burleigh are provided at no charge; please check on specific
models.
-
Line Frequency Noise - The most common problem with any SEM used for
lithography is line frequency noise from the local environment. When present,
the noise can almost always be seen in the imaging at high magnification,
however, to the untrained eye the distortion from the noise can be difficult
to identify. Another way to observe such noise is to put the SEM into a "Spot
Mode" with the beam hitting an area where the sample image rapidly changes
from bright to dark. If the beam is oscillating on the sample due to line
frequency pickup, the image output signal will display a line frequency
oscillation when viewed on an oscilloscope. Rotate the sample to check for
beam movement in different directions. If line frequency noise is observed,
additional information can be found at
"Solutions to Problems:
Pattern Noise from Environment". It may also be useful to do this "Spot
Mode" test with the beam hitting a completely uniform area. In this case,
any oscillation seen in the image output signal will not be from movement
of the beam, but rather from noise pickup in the imaging circuitry, or possibly
variations in the beam current itself.
-
XY Inputs - The SEM must have XY inputs if it is to be used for
lithography. Check the SEM documentation and find the XY input connector(s).
There will also be some method to select the external inputs for beam control
instead of the default internal scan generator. Often, this is an input which
is switched between 0v and 5v or between an open and closed connection. By
connecting a voltage supply to the XY inputs and enabling them, you will
often be able to see the beam move on the CRT in response to the voltage
supplied. For an analog SEM, before enabling the XY inputs, turn down
the CRT and/or SEM brightness to avoid burning the CRT screen. On SEMs
with digital image display, the CRT will typically not display the spot in
external mode. In this case, the image output signal on the SEM can be monitored
while changing the XY voltages to move the beam from a dark region to a bright
region on the sample. The change in the image signal should confirm that
the beam is actually moving. If the SEM has or has had an x-ray mapping system
or an active digital imaging system attached, then it must also have XY inputs
for beam control, although if there is only a single XY input, an input relay
adapter will be necessary to simultaneously connect NPGS and the other accessory.
|
Home |
Overview | Questions
& Answers | System Description
| Microscope Considerations
| User List |
Sample Pictures |
Other Resources |
Book List | User
Notices | Send
E-mail
|